01-ika-0008802500
01 ika 0008802500
02 20240426141108

IKA C 200

Varenr. IKA0008802500

C 200-System consists of the following components:
Measuring cell C 200
Decomposition vessel C 5010
Oxygen filling station C 248
Consumables for calibrations and installation

Beskrivelse

C 200

C 200-System consists of the following components:
Measuring cell C 200
Decomposition vessel C 5010
Oxygen filling station C 248
Consumables for calibrations and installation

Space-saving and low cost combustion calorimeter for determing gross calorific values of liquid and solid samples.

Validation according to DIN 51900,ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711.

Suitable for teaching and training, as well (f.e tecnical- schools, universities) and for industriell laboratories with less analyses..

Manual handling for water and oxygen filling.

Four different working methods, isoperibolic, dynamic, manual, time controlled makes possible an ideal adjustment of different operating tasks.

Clear and easily explained display which is simply to use .

With an external power pack, "world tension useable" from 100 - 240 V AC, 50/60 Hz. The operating voltage of the calorimeter is 24 V DC low- voltage.

Friendly in aftersales

Manual ( teaching mode): ignition and end of the measurement will be done by the operator himself; the temperature increase will be indicated at the display each minute.
All calculations have to be done manually.
At the other three modes ignition and calculation of the gross calorific value will be done automatically.
The gross calorific value will be shown on the display. Acid corrections and calculations of the net calorific value have to be done manually.

The modes have different measuring times:
Isoperibolic: approx: 17 min.
Dynamic : approx. 8 min
Manual : approx. 17 min
(depend on the operator)
Time controlled : 14 min

The decomposition vessel can be equiped in to use an combustible crucibles C 14
(accessory C 5010.4 attachment is neessary)

User-friendly software C 6040 Calwin for controlling the calorimeter and administration of measuring data (asseccory). Transfer of the datas to Excel or Word are possible
Up to eight IKA measuring cells can be controlled by a single PC, using a multiserial plug in card PCI 8.2 (accessory)

Scope of delivery

  • C 200
  • C 5010 Decomposition vessel, standard
  • C 248 Oxygen filling station
  • Ignation adapter
  • C 200.1 Measuring cup 2000 ml
  • Water emptying hose

Technical Data

Measuring range max.40000 J
Measuring mode dynamic 25°Cyes
Measuring mode isoperibol 25°Cyes
Measuring time dynamic approx.8 min
Measuring time isoperibol approx.17 min
Reproducibility dynamic (1g benzoic acid NBS39i)0.1 %RSD
Reproducibility isoperibol (1g benzoic acid NBS39i)0.1 %RSD
Working temperature max.25 °C
Temperature measurement resolution0.0001 K
Interface printerCentronix
Interface PCRS232
Decomposition vessel C 5010yes
Works according to DIN 51900yes
Works according to DIN EN ISO 1716yes
Works according to DIN EN ISO 18125yes
Works according to DIN EN 15400yes
Works according to DIN CEN TS 14918yes
Works according to DIN CEN/TS 16023yes
Works according to DIN SPEC 19524yes
Works according to ASTM D240yes
Works according to ASTM D4809yes
Works according to ASTM D5468yes
Works according to ASTM D5865yes
Works according to ISO 1928yes
Works according to GOST Certifiedyes
Dimensions (W x H x D)400 x 400 x 400 mm
Weight28.912 kg
Permissible ambient temperature20 - 25 °C
Permissible relative humidity80 %
Protection class according to DIN EN 60529IP 20
RS 232 interfaceyes
Voltage100 - 240 V
Frequency50/60 Hz
Power input120 W
Specifikationer
Varenr. IKA0008802500
Antal / pk 1
Enhed stk
Min. bestilling 1
CE-godkendt Ja